Question
if a sound beam moves from air into soft tissue at an angle of 60° what is the angle of reflection?
Question
You switch from a 5 MHz curved probe to a 2.5 MHz sector probe to image deeper in the abdomen how is the wavelength affected by the change to the new probe
Question
even and odd harmonic frequencies are created by
Question
if more cycles are added to a pulse which of the following resolution is degraded
Question
why can’t ultrasound probes be thermally sterilized
Question
The doppler tracing of the flow in a vessel 1 cm distal to a stenosis will demonstrate ___ than at the stenosis
Question
when scanning using harmonic imaging the second harmonic frequency is 10 MHz. what is the fundamental frequency
Question
when transmural pressure in a vein is high
Question
for each transducer shown there is an U/S image produced. when comparing the two, which of the following statements is true regarding the image produced by the transducer on the right ?
Question
which of the following should be increased when evaluating venous flow on doppler
SPI version D pt 2
Score
0 / 10
Correct rate
0%
Correct: 0
Incorrect: 0
Unanswered: 10
Question list
1
if a sound beam moves from air into soft tissue at an angle of 60° what is the angle of reflection?C) 60°
2
You switch from a 5 MHz curved probe to a 2.5 MHz sector probe to image deeper in the abdomen how is the wavelength affected by the change to the new probeC) doubles
3
even and odd harmonic frequencies are created byD) non linear sound wave propagation
4
if more cycles are added to a pulse which of the following resolution is degradedC) axial
5
why can’t ultrasound probes be thermally sterilizedC) if the probe is heated above the curie point the piezoelectric properties will be lost
6
The doppler tracing of the flow in a vessel 1 cm distal to a stenosis will demonstrate ___ than at the stenosisA) greater bandwidth
7
when scanning using harmonic imaging the second harmonic frequency is 10 MHz. what is the fundamental frequencyB) 5 Mhz
8
when transmural pressure in a vein is highD) the vein becomes enlarged and rounded
9
for each transducer shown there is an U/S image produced. when comparing the two, which of the following statements is true regarding the image produced by the transducer on the right ?B) the image will show improved spatial resolution
10
which of the following should be increased when evaluating venous flow on dopplerA) color priority
11
which of the following correctly describes an amplifierA) each returning reflection is made stronger/ larger for additional processing and storage
12
what system parameter is being assessed on the phantom imageD) depth calibration
13
the primary advantage of continuous wave doppler isD) absence of the Nyquist limit
14
which of the following exams does not require angle corrected doppler evaluationD) IVC duplex
15
if the transmitted frequency is 5 MHz and the reflected frequency is 4.9975 MHz how will the doppler shift be displayedD) under the baseline
16
which of the following is a complication of series stenosis in an arteryD) velocity and stenosis can be underestimated at the second stenosis
17
You noticed the cord of the echo transducer is slightly separated from the back of the probe which of the following describes the appropriate action to take nextD) place a service call for the probe and reschedule all patients until it is repaired
18
If you use a tissue phantom that contains a substance with an average propagation speed of 1600 cm/s to evaluate a standard ultrasound system, how this affect the depth calibration testingC) reflectors will be more shallow than they actually are located within the phantom
19
which of the following correctly, describes the negative predictive value (NPV) of a testing techniqueC) if the test is negative NPV is the probability that the patient does not have the disease
20
what type of precautions is required for a patient with MRSAD) contact precautions
21
If the Space between the dotted lines indicates a total time of 0.5 seconds what is the frequency of the sound transmission?A) 4 Hz
22
Which of the following ultrasound machine controls affects the power of the beamB) transmit gain
23
Potential bioeffects of ultrasound can be minimized by which of the followingC) minimizing dwell time
24
According to the attached color map, what color will demonstrate for laminar flow that is moving away from the probe at 35 cm/sB) blue
25
Which system control will simultaneously enhanced far field echoes and diminished near field echoesC) time gain compensation
26
The use of multiple focal zones will improve ___ but degrade ___C) lateral resolution, temporal resolution
27
Which of the following is not necessary in construction of a cavitated micromachine ultrasound transducerA) matching layer
28
Curved PE elements and acoustic lenses will improve which two types of resolutionA) elevational and lateral
29
Which of the following should be activated to create a smoother appearance to the imageD) persistence
30
Patient can revoke consent from any invasive procedureA) at any time for any reason
31
Blooming artifacts occursD) when color obscures the gray scale display of soft plaque because the resolution of color Doppler is lower than the gray scale
32
What part of the ultrasound system is responsible for changing the element function back-and-forth from transmit to receiveB) beam former
33
Increasing the rejection or suppression setting can be used toA) increase the contrast of the image to produce a more black-and-white image
34
The entire image is too bright. What console adjustments should you make firstD) decrease the output power
35
What system settings should be adjusted to better visualize a possible stone in the ureter posterior to the bladderA) turn off spatial compounding
36
If the PRF is set to the maximum level, how can the artifact be correctedD) utilize CW
37
What artifact is produced by the interference of reflected waves from many tiny reflectorsD) speckle
38
As the transducer frequency increases the quality factor ___D) increases
39
Which of the following is not part of the sonographer responsibilities for an ultrasound exam?B) offering advice on ways to alleviate symptoms until the doctor treats the problem treats the problem
40
Lower the Tx frequency willA) increase the mechanical index
41
Which of the following will improve signal to noise ratio of the imageB) increasing output power
42
What is the best way to determine if a displayed structure is real or an image artifact when scanningD) use multiple imaging windows to evaluate the area
43
Which of the following describes a vector array transducerD) has a sector shaped field of view with a flat top
44
What type of probe is preferred for cardiac imagingB) phased sector
45
Which of the following describes power DopplerD) color is assigned to the Doppler shift based on the number of moving red blood cells
46
____ is the term used to describe the concentric layers of blood flow within a normal vesselD) laminar flow
47
Harmonic imaging leads toC) decreased image clutter
48
A linear transducer providing a trapezoidal image demonstrates degraded ___ resolution in the far field of the imageB) lateral
49
Which of the following will increase the duty factor of the pulse ultrasound probeA) decreasing the image depth
50
Which of the following will increase the amount of color displayed in a vessel located deep in the bodyD) increased frame rate
51
What console setting can be changed to adjust the propagation speed of the transmitted waveD) none of the above
52
Which of the following is the most likely explanation for findings displayed on the imageA) the deeper vessel is a mirror image artifact caused by the Apex of lung
53
Steering the color sample box will cause changes in theA) Doppler shift and angle of insonation
54
Adjust you to the following parameters will correct the Doppler artifact displayedC) change to lower frequency probe
55
What is the best way to visualize a suspected mass that is isoechoic to surrounding tissues?A) minimize the dynamic range to 60-65 dB
56
which of the following is true regarding continuous wave transducer structure and functionA) CW probes do not require any backing material
57
if the data line indicates a time of one second from the start of wave transmission, what is the frequency of the sound transmissionD) 4 Hz
58
as a soundwave travels through soft tissue what happens to the period of the waveC) remains constant
59
spatial resolution is:D) the minimum reflector separation required to produce separate echoes
60
the coupling gel use for ultrasound exams provides the same function as what transducer componentC) matching layer
61
____ is defined as the ratio of two sound intensitiesC) decibel