Question
contribute to overall radiographic quality
Question
2 Geometric Qualities ______ - distinct representation of objects true borders, edges ______ - misinterpretation of true size or shape of an object
Question
degree of difference or ratio between
Question
no differences in density
Question
object with different absorption characteristics produces an image with varying densities
Question
Evaluating radiographic quality in contrast is ______
Question
- controlling factor of contrast - difference in potential applied to x-ray tube - directly proportional to average x-ray spectrum or x-ray quality - plays role in adjusting amount of penetration in an acquisition - energy of x-rays (higher penetrability)
Question
examples of: low subject contrast > high kVp > long scale contrast
Question
examples of: high subject contrast > low kVp > short scale contrast >
Question
properties that comprise visibility of detail. Appear Gray
POI P2
Score
0 / 10
Correct rate
0%
Correct: 0
Incorrect: 0
Unanswered: 10
Question list
1
contribute to overall radiographic qualityGeometric Qualities
2
2 Geometric Qualities ______ - distinct representation of objects true borders, edges ______ - misinterpretation of true size or shape of an object- Recorded detail - Distortion
3
degree of difference or ratio betweenRadiographic Contrast
4
no differences in densityHomogenous Object
5
object with different absorption characteristics produces an image with varying densitiesDifferential Absorption
6
Evaluating radiographic quality in contrast is ______Subjective
7
- controlling factor of contrast - difference in potential applied to x-ray tube - directly proportional to average x-ray spectrum or x-ray quality - plays role in adjusting amount of penetration in an acquisition - energy of x-rays (higher penetrability)kVp
8
examples of: low subject contrast > high kVp > long scale contrastAir/Fluid
9
examples of: high subject contrast > low kVp > short scale contrast >Bone
10
properties that comprise visibility of detail. Appear GrayContrast
11
visible to human eyeVisibility of detail
12
difference between adjacent densitiesImage Contrast
13
number of useful visible densities or shades of grayScale of Contrast
14
described by scale of contrastRadiographic Images
15
conversion factor between exposure to ionizing radiationTechnique Conversion Factors
16
what to do To maintain exposure when increasing kVp:- increase kVp by 15% - reduce mAs to half
17
what to do To maintain exposure when decreasing kVp:- decrease kVp 15% - multiply mAs by 2
18
tissues that makeup the human body that attenuate the beam of radiationPatient Factors
19
- device used to improve contrast of radiographic image - absorbs scattered radiation before it reaches image receptor - employed when body part thickness is greater than 10cm or kVp is 60 - used for obese patientsGrid
20
3 X-ray Interaction through the body PBI- Pass through the body - Be absorbed by body - Interact & change direction
21
interaction that produces scatter radiationCompton Interaction
22
3 Amount of Scatter Radiation increase with: ILD- Increased patient thickness - Larger field sizes - Decrease in atomic number of tissue
23
account for lead content of gridGrid Conversion Factor aka Bucky Grid
24
- placed in cassette close contact with film, function is to reinforce the action of x-rays - amplify the incoming x-ray beam - introduced by Thomas Edison in 1896Intensifying Screens
25
2 TYPES OF INTENSIFYING SCREENS- Calcium Tungstate: blue light - Rare Earth: green light
26
percent of x-rays that reach cassette33%
27
invisible product of x-ray filmLatent Image
28
4 PARTS OF INTENSIFYING SCREEN BRPP- Base Layer - Reflective Layer - Phosphor Layer - Protective Coat
29
layer of IS that is polyester or cardboard or metalBase Layer
30
layer of IS that is made up of magnesium oxide or titanium dioxideReflective Layer
31
active layer of IS , absorbs energyPhosphor Layer
32
protects phosphor layer from abrasionsProtective Coat
33
3 Types of Beam Restrictors ACC- Aperture Diaphragms - Cones/Cylinder - Collimators
34
beam restrictor that is a flat sheet of metalAperture diaphragms
35
circular aperture diaphragmsCones/Cylinder
36
commonly employed beam restrictionCollimators
37
purpose is to deposit enough metallic silver at latent image sitesFilm Processing
38
4 Processing Steps> Developing - silver is deposited at latent image > Fixing - clearing agent > Washing - water to remove developer solutions > Drying - hot air
39
critical part of archiving processStorage
40
radiograph time or expiry5-7 years
41
- visibility of anatomy and sign of pathology - affected by photographic and geometric qualityImage Quality