RPC F1

36 questions • 2024-05-31
E
Report

Question

Have several causes, most of which are related to the thermal characteristics of the x-ray tube

Answer

Question

must be dissipated for the x-ray tube to continue to function.

Answer

Question

Three Methods of Heat Dissipation

Answer

Question

results in reduced tube life (also the first reason for tube failure)

Answer

Question

caused by excessive temperature of the anode

Answer

Question

what may result to cracking of the anode.

Answer

Question

should not be applied to a cold cathode.

Answer

Question

results from maintaining the anode at elevated temperatures for prolonged times.

Answer

Question

what’s the final cause of tube failure?

Answer

Question

caused when vaporized tungsten coats the inside of the glass or metal enclosure is interacted with electrons.

Answer

RPC F1

TestMaker

Score

0 / 10

Correct rate

0%

Correct: 0

Incorrect: 0

Unanswered: 10

 
ankimaker.com

Question list

  • 1

    Have several causes, most of which are related to the thermal characteristics of the x-ray tube

    X-Ray Tube Failure

  • 2

    must be dissipated for the x-ray tube to continue to function.

    Heat

  • 3

    Three Methods of Heat Dissipation

    Radiation - transfer of heat by emission of infrared radiation. Convection - transfer of heat by the movement of a heated substance from one place to another Conduction - transfer of heat from one area of an object to another.

  • 4

    results in reduced tube life (also the first reason for tube failure)

    Excessive heat

  • 5

    caused by excessive temperature of the anode

    Localized Surface Melting & Pitting

  • 6

    what may result to cracking of the anode.

    Too rapid increase in temperature of the anode

  • 7

    should not be applied to a cold cathode.

    Maximum radiographic technique

  • 8

    results from maintaining the anode at elevated temperatures for prolonged times.

    Second type of tube failure

  • 9

    what’s the final cause of tube failure?

    Tube Arcing

  • 10

    caused when vaporized tungsten coats the inside of the glass or metal enclosure is interacted with electrons.

    Tube Arcing

  • 11

    - causes disturbed electrical balance on the x-ray tube, - causing abrupt, intermittent changes in tube current

    Tube Arcing

  • 12

    guides the the radiologic technologists in using the x-ray tube

    Rating Chart

  • 13

    Three types of Rating charts

    - Radiographic Rating Chart - Anode Cooling Chart - Housing Cooling Chart

  • 14

    most important of the three types of rating charts

    Radiographic Rating Chart

  • 15

    shows scales of mA and kVp

    X and Y

  • 16

    - contains the thermal capacity of the anode and its heat dissipation characteristics. - has a limited capacity in storing heat.

    Anode Cooling Chart

  • 17

    - has a shape similar to that of the anode cooling chart and is used precisely the same way. - complete cooling after maximum heat capacity requires from 1 to 2 hours

    Housing Cooling Chart

  • 18

    - describes the energy deposition of a particular type of radiation, which largely determine the biological consequence of radiation

    Linear Energy Transfer

  • 19

    - proportional to the square of the charge of the particle. - inversely proportional to the particle's kinetic energy.

    Linear Energy Transfer

  • 20

    what is the unit of LET

    keV/um

  • 21

    High Linear Energy Transfer Radiation range

    3-200 keV/um

  • 22

    Low Linear Energy Transfer Radiation

    0.2-3 keV/um

  • 23

    High Linear Energy Transfer is mediated by what?

    alpha particles, proton, neutrons

  • 24

    - Greater density of interactions at cellular level - produce biological damage in a given volume of tissue

    High Linear Energy Transfer

  • 25

    Low Linear Energy Transfer is mediated by what?

    Electrons, Positrons, Gamma rays, X-rays

  • 26

    less likely to produce tissue damage in the same volume of tissue

    Low Linear Energy Transfer

  • 27

    conveys which radiographic techniques are safe and unsafe for x-ray tube operation

    Radiographic Rating Charts

  • 28

    average energy deposited per unit path length along the track of an ionizing particle

    Linear Energy Transfer

  • 29

    LET of Linac X-rays (6-15 MeV)

    0.3 keV/um

  • 30

    LET of Beta Particle (1 MeV)

    0.3 keV/um

  • 31

    LET of Cobalt-60 Y-rays

    0.2 keV/um

  • 32

    LET of 250 kVp x-rays (standard)

    2 keV/um

  • 33

    LET of 150 MeV protons

    0.5 keV/um

  • 34

    LET of Neutrons

    0.5-100 keV/um

  • 35

    LET of Alpha Particles

    50-200 keV/um

  • 36

    LET of Carbon ions

    40-90 keV/um