Question
The fluoroscopic exposure control should be of the ____ ____ type
Question
Source to ____ Distance SSD must NOT BE LESS than __ cm on stationary fluoroscopes SSD must NOT BE LESS than __ cm on mobile fluoroscopes. Increased SSD: reduces entrance ____ exposure
Question
serves as a primary protective barrier and must be 2 mm Pb equivalent
Question
_____ - 2.5mm Al ___ should be measured when filtration is unknown
Question
unexposed border should be visible at all heights above the tabletop
Question
unexposed border is visible on the image monitor (35 cm above the tabletop)
Question
moved at the end of the table during fluoroscopy leaving a 5 cm opening in the side of the table and Covered with ____ mm Pb
Question
It should be position between fluoroscopist & patient Equivalent: 0.25 mm Pb
Question
- produces an audible sound when the fluoroscopic time has exceeded _ minutes - Designed to ensure that the radiologist is aware of the relative beam-on time during each procedure
Question
The intensity of the x-ray beam at the tabletop of a fluoroscopy should not exceed ___ mGy/min for each mA of operation at ___ kVp
RADPRO 3
Score
0 / 10
Correct rate
0%
Correct: 0
Incorrect: 0
Unanswered: 10
Question list
1
The fluoroscopic exposure control should be of the ____ ____ typedead man
2
Source to ____ Distance SSD must NOT BE LESS than __ cm on stationary fluoroscopes SSD must NOT BE LESS than __ cm on mobile fluoroscopes. Increased SSD: reduces entrance ____ exposure1. Skin 2. 38 3. 30 4. skin
3
serves as a primary protective barrier and must be 2 mm Pb equivalentFluoroscopic image receptor assembly
4
_____ - 2.5mm Al ___ should be measured when filtration is unknown1. Filtration 2. HVL
5
unexposed border should be visible at all heights above the tabletopAutomatic Collimators
6
unexposed border is visible on the image monitor (35 cm above the tabletop)Rationale
7
moved at the end of the table during fluoroscopy leaving a 5 cm opening in the side of the table and Covered with ____ mm PbBucky Tray 0.25
8
It should be position between fluoroscopist & patient Equivalent: 0.25 mm PbProtective Curtain
9
- produces an audible sound when the fluoroscopic time has exceeded _ minutes - Designed to ensure that the radiologist is aware of the relative beam-on time during each procedureCumulative Timer - 5 minutes
10
The intensity of the x-ray beam at the tabletop of a fluoroscopy should not exceed ___ mGy/min for each mA of operation at ___ kVp2.1 80
11
If an optional high-level control: ___ mGy/min If there is no optional high-level control: ___ mGy/min during fluoroscopy.1. 200 2. 100
12
________ _________ receive most of their occupational radiation exposure during _________- Radiologic technologists - Fluoroscopy
13
where lead bonded to sheet rock or wood panelling is used inPrimary Barrier
14
results when the useful beam intercepts any objects causing some x-rays to be scatteredScatter Radiation
15
Radiation emitted from the x-ray tube housing in all directions other than that of the useful beam Limit: ___ mR/hr at 1 m1. Leakage Radiation 2. 100
16
depends on the distance between the source of radiation & the barrierDistance
17
the use of the area that is being protectedOccupancy
18
Under Control: _________ _____: an area occupied primarily by radiology personnel & patients Limits: less than ___ mrem/week or 1 mSv/week1. Controlled Area 2. 100
19
product of the maximum mA & the number of x-ray examinations performed per week (expressed in mAmin/week)Workload
20
The percentage of time during which the x-ray beam is on & directed toward a particular protective barrierUse Factor
21
___: It is used as a measure of penetrability General Radiography: ___ kVp Mammography: ___ kVp1. kVp 2. 100 3. 30
22
6 Factors that affect Barrier Thickness DOCWUK> Distance > Occupancy > Control > Workload > Use Factor > kVp
23
2 Alternative for lead- Concrete or Concrete Block - Brick
24
4 inches of masonry is equivalent to ______ of lead.1/16 inch
25
____ of masonry is equivalent to 1/16 inch of lead.4 inches