The fluoroscopic exposure control should be of the ____ ____ type

Answer

Question

Source to ____ Distance SSD must NOT BE LESS than __ cm on stationary fluoroscopes SSD must NOT BE LESS than __ cm on mobile fluoroscopes. Increased SSD: reduces entrance ____ exposure

Answer

Question

serves as a primary protective barrier and must be 2 mm Pb equivalent

Answer

Question

_____ - 2.5mm Al ___ should be measured when filtration is unknown

Answer

Question

unexposed border should be visible at all heights above the tabletop

Answer

Question

unexposed border is visible on the image monitor (35 cm above the tabletop)

Answer

Question

moved at the end of the table during fluoroscopy leaving a 5 cm opening in the side of the table and Covered with ____ mm Pb

Answer

Question

It should be position between fluoroscopist & patient Equivalent: 0.25 mm Pb

Answer

Question

- produces an audible sound when the fluoroscopic time has exceeded _ minutes - Designed to ensure that the radiologist is aware of the relative beam-on time during each procedure

Answer

Question

The intensity of the x-ray beam at the tabletop of a fluoroscopy should not exceed ___ mGy/min for each mA of operation at ___ kVp

Answer

RADPRO 3

TestMaker

Score

0 / 10

Correct rate

0%

Correct: 0

Incorrect: 0

Unanswered: 10

 
ankimaker.com

Question list

  • 1

    The fluoroscopic exposure control should be of the ____ ____ type

    dead man

  • 2

    Source to ____ Distance SSD must NOT BE LESS than __ cm on stationary fluoroscopes SSD must NOT BE LESS than __ cm on mobile fluoroscopes. Increased SSD: reduces entrance ____ exposure

    1. Skin 2. 38 3. 30 4. skin

  • 3

    serves as a primary protective barrier and must be 2 mm Pb equivalent

    Fluoroscopic image receptor assembly

  • 4

    _____ - 2.5mm Al ___ should be measured when filtration is unknown

    1. Filtration 2. HVL

  • 5

    unexposed border should be visible at all heights above the tabletop

    Automatic Collimators

  • 6

    unexposed border is visible on the image monitor (35 cm above the tabletop)

    Rationale

  • 7

    moved at the end of the table during fluoroscopy leaving a 5 cm opening in the side of the table and Covered with ____ mm Pb

    Bucky Tray 0.25

  • 8

    It should be position between fluoroscopist & patient Equivalent: 0.25 mm Pb

    Protective Curtain

  • 9

    - produces an audible sound when the fluoroscopic time has exceeded _ minutes - Designed to ensure that the radiologist is aware of the relative beam-on time during each procedure

    Cumulative Timer - 5 minutes

  • 10

    The intensity of the x-ray beam at the tabletop of a fluoroscopy should not exceed ___ mGy/min for each mA of operation at ___ kVp

    2.1 80

  • 11

    If an optional high-level control: ___ mGy/min If there is no optional high-level control: ___ mGy/min during fluoroscopy.

    1. 200 2. 100

  • 12

    ________ _________ receive most of their occupational radiation exposure during _________

    - Radiologic technologists - Fluoroscopy

  • 13

    where lead bonded to sheet rock or wood panelling is used in

    Primary Barrier

  • 14

    results when the useful beam intercepts any objects causing some x-rays to be scattered

    Scatter Radiation

  • 15

    Radiation emitted from the x-ray tube housing in all directions other than that of the useful beam Limit: ___ mR/hr at 1 m

    1. Leakage Radiation 2. 100

  • 16

    depends on the distance between the source of radiation & the barrier

    Distance

  • 17

    the use of the area that is being protected

    Occupancy

  • 18

    Under Control: _________ _____: an area occupied primarily by radiology personnel & patients Limits: less than ___ mrem/week or 1 mSv/week

    1. Controlled Area 2. 100

  • 19

    product of the maximum mA & the number of x-ray examinations performed per week (expressed in mAmin/week)

    Workload

  • 20

    The percentage of time during which the x-ray beam is on & directed toward a particular protective barrier

    Use Factor

  • 21

    ___: It is used as a measure of penetrability General Radiography: ___ kVp Mammography: ___ kVp

    1. kVp 2. 100 3. 30

  • 22

    6 Factors that affect Barrier Thickness DOCWUK

    > Distance > Occupancy > Control > Workload > Use Factor > kVp

  • 23

    2 Alternative for lead

    - Concrete or Concrete Block - Brick

  • 24

    4 inches of masonry is equivalent to ______ of lead.

    1/16 inch

  • 25

    ____ of masonry is equivalent to 1/16 inch of lead.

    4 inches