記憶度
15問
35問
0問
0問
0問
アカウント登録して、解答結果を保存しよう
問題一覧
1
Luminous intensity
Candela
2
Electric current
Ampere
3
Thermodynamic temperature
Kelvin
4
Electrical potential
Volt
5
Temperature
Degree Celcius
6
Arrange ascending order
zepto, atto, femto, pico, nano, micro, milli, centi, deci, deka, hecto, kilo, mega, giga, giga, tera, peta, exa
7
Number of moles of solute per liter of solution
molarity
8
Number of moles of solute per kilogram solvent
molality
9
Number of equivalent weights of solute per liter of solution
normality
10
What is a valence?
Either
11
Percent solution can be expressed as the following, except;
% v/w
12
Expression of relative concentration
dilution
13
Done when the analyte concentration is too high and exceeds the detection limit of the method
dilution of sample
14
Done when a weaker reagent concentration is needed than the stock substance available
Dilution of stock solution
15
Amount of something relative to another; expressed as part per part or part per whole
ratio
16
Average or arithmetic mean
mean
17
Midpoint of a data set after the values have been ranked-ordered
median
18
Most frequently occurring value in a data set
mode
19
It is the difference between the highest and lowest values; greatly affected by outlier.
range
20
The distribution of data points around the mean
standard deviation
21
The best indicator of precision
coefficient of variation
22
What is the usual confidence interval of a guassian/normal distribution?
95%
23
Verifying a reference interval requires how many study? individuals
atleast 20
24
RI is adopted if ____ of the subjects fall outside the range
<10%
25
Establishing a reference interval requires how many study individuals?
atleast 120
26
What CI is the RI based?
95%
27
Compares the means of two groups of data or the accuracy of two methods
t-test
28
Compares the standard deviations of two groups of data or the precision of two procedures
f test
29
It is used to compare two methods using the best fit line through the data points
linear regression
30
The x-axis is the dependent variable/new method. The Y-axis is the independent/reference method.
Both statements are false
31
Ability of a method to detect the smallest concentration of an analyte
analytical sensitivity
32
Ability of a method to detect only the analyte of interest
analytical specificity
33
Ability of a test to detect a given disease or condition; proportion of individuals with the disease who have a positive test result
diagnostic sensitivity
34
Ability of a test to detect the absence of a given disease or condition; proportion of individuals with no disease who have a negative test result
diagnostic specificity
35
Used to compare results obtain on a high and low control serum from different laboratories
Youden Twin plot
36
Precision study is usually done by running
two control materials twice a day over a 10 day period
37
It is the most widely used approach to quality improvement in the healthcare
PDCA cycle
38
Provides both a management philosophy or organizational development and a management process for improvement of quality in all aspects of work
TQM (Total Quality Management)
39
Based on statistics and quantitative measurements applied in the clinical laboratory to reduce the frequency of test errors or the number of DPMOs
Six Sigma
40
It is designed to reduce waste, increase efficiency and improve customer satisfaction; often used in clinical laboratories to improve TAT.
Lean
41
Categories of wastes:
Defects, Overproduction, Wasting , Non-utilised talent, Transport , Inventory, Motion, Excess processing
42
Understands the basics of lean six sigma
yellow belt
43
Team members who contribute approximately 20% of their time to QI projects while delivering their normal job functions
green belt
44
Team leaders who dedicate as much as 100% of their time to QI projects
black belt
45
Black belt with at least two years of experience; lean six sigma advisors/coaches
master
46
Pre-analytical includes;
Test requisition, Patient preparation , Patient identification , Specimen collection , Labeling, Specimen transport, Specimen reception, Handling, Preparation
47
Analytical includes;
Reagents, Preventive maintenance of equipment , Calinration, Analysis of samples, Quality control
48
Post-analytical includes;
Verification of calculations and reference ranges, Flagging and notification of panic/critical values, Delta checks, Reporting of results, Interpretation
49
Inappropriate test requested
Pre-analytical
50
Handwriting not legible
Pre-analytical
51
Special requirements not specified
Pre-analytical
52
Delayed order
Pre-analytical
53
Incorrect tube or container
Pre-analytical
54
Incorrect patient identification
Pre-analytical
55
Inadequate specimen volume
Pre-analytical
56
Invalid specimen
Pre-analytical
57
Wrong collection time
Pre-analytical
58
Improper transport conditions
Pre-analytical
59
Instrument not calibrated correctly
Analytical
60
Instrument precision problem
Analytical
61
Specimen mix up
Analytical
62
Incorrect sample volume
Analytical
63
Interfering substance present
Analytical
64
Poorly written procedure
Analytical
65
Deteriorating reagent
Analytical
66
Contaminated control solutions
Analytical
67
Calibration error
Analytical
68
Report not posted in chart, not legible, or delayed
Post-analytical
69
Transcription error
Post-analytical
70
Interfering substances not recognized
Post-analytical
71
Specificity of test not understood
Post-analytical
72
Precision limitations not recognized
Post-analytical
73
Analytical sensitivity not appropriate
Post-analytical
74
Previous values not available for comparison
Post-analytical
75
Ability to maintain accuracy and precision over an extended period of time during which equipment, reagents, and personnel may change
Reliability
76
Closeness of the result to the true or actual value
accuracy
77
Ability to produce a series of results that agree closely with each other; commonly expressed in terms of coefficient of variation; also called repeatability or reproducibility
Precision
78
Internal Quality Control involves the analysis of at least __ levels of control every ___
2; 24 hours
79
In internal quality control, initial control limits are established by analyzing controls for at least ___ consecutive days or runs
20
80
Involves testing blind samples sent periodically by regulatory agencies to participating laboratories
external quality assessment
81
Due to chance or an unpredictable cause; does not recur in regular pattern
Random error
82
Usually a one time error and represents imprecision
Random error
83
Random error
All of the choices
84
Warning!!!
1 2s
85
Causes random errors
instrument instability such as temperature and voltage fluctuations, dirty glasswares, sampling error, improper mixing of sample and reagent, operator variability such as pipeting error, anticoagulant or drug interference
86
Causes systematic errors
deterioration of reagents, change in reagent lot, unstable reagent blanks, calibration error, changes in standard concentration, contaminated control solutions, diminishing lamp power, dirty photo meter, faulty ISE, pipettor maintenance error
87
A recurring error inherent in test procedure; influences observations in one direction
Systematic error
88
Maybe constant or proportional
Systematic error
89
Systematic error
2 2s, 4 1s, 8 1s, 10 x
90
It is an abrupt change in the distribution of controlled values such that they accumulate on one side of the mean for six consecutive days
Shift
91
Major cause of shift
calibration error or error in standard preparation
92
Major cause of trend
deterioration of reagents
93
Gradual change in the distribution of control values such that they continue to increase or decrease over a period of six consecutive days passing through the mean
trend
94
Highly deviating control values
outliers
95
Prepared by plotting the algebraic sum of the difference between each quality control result and the mean on the y-axis and the run number on the x-axis
CUSUM chart
96
Used for interlaboratory comparison of monthly means by plotting the mean value for one sample on the y-axis and the mean value for another on the x-axis
Tonks-Youden plot
97
It is an effective method for comparing both within-laboratory and between-laboratory variability
Tonks-Youden plot