暗記メーカー
ログイン
SPI version A
  • Karla Alvarez

  • 問題数 100 • 6/7/2024

    記憶度

    完璧

    15

    覚えた

    35

    うろ覚え

    0

    苦手

    0

    未解答

    0

    アカウント登録して、解答結果を保存しよう

    問題一覧

  • 1

    which of the following is displayed with spectral Doppler but not with color Doppler

    A) peak flow velocity

  • 2

    The cine function on the ultrasound system

    B) is a type of post processing

  • 3

    What type of precautions is required for a patient with MRSA

    D) contact precautions

  • 4

    ____ is defined as the time from onset of systole to the peak systole occurs

    A) acceleration time

  • 5

    if a renal Doppler exam demonstrates stenosis, but normal flow is identified on angiography evaluation the results are described as

    C) false positive

  • 6

    increasing the number of bits per pixel in an image memory will improve the

    C) contrast resolution

  • 7

    what changes occurred to the ultrasound beam when time pulses are emitted from left to right from the elements of the phased transducer

    D) the beam is steered to the right

  • 8

    as a Doppler incident angle increases to 90° the detected Doppler shift varies and approaches what value

    B) 0 kHz

  • 9

    multiply density of the tissue by propagation speed in the tissue will provide a value for what tissue characteristic

    B) acoustic impedance

  • 10

    which of the following is a limitation of analog Doppler?

    D) unable to detect post stenotic turbulence and spectral broadening

  • 11

    The acoustic impedance of the matching layer

    C) is usually very low

  • 12

    ____ is the predominant factor causing attenuation of the ultrasound beam and this can be counteracted by ____

    A) sound absorption, increasing the TGC settings for the far field

  • 13

    Assuming line density is constant at two lines per degree and you change from a 90° sector to a 45° sector how many lines will now be produced

    B) 90

  • 14

    which of the following is a limitation of power Doppler

    C) has slower frame rates than color doppler

  • 15

    which of the following statements is true regarding the image produced by the transducer on the left?

    D) the frame rate will be higher for the image

  • 16

    what color Doppler artifact is demonstrated on the image

    A) flash

  • 17

    increasing the number of pulses in a color packet will

    D) increase frame rate

  • 18

    according to the attached color map, what color would demonstrate for blood that is moving toward the probe at an average velocity of 100cm/s

    B) white/ light blue

  • 19

    which of the following sound/ tissue interactions does not cause redirection of the incident wave

    A) absorption

  • 20

    if two calcifications are located side-by-side in the liver with a 3 mm distance between them and a beam width of 6 mm how will the structure appear on the image?

    A) as a single elongated horizontal reflection

  • 21

    tissue Doppler imaging uses ____ settings than PW doppler

    D) lower filter

  • 22

    If the entire image is too dark, what console adjustment should you make first?

    A) increase the overall gain

  • 23

    What should have been done to adjust and improve the color Doppler display on the image

    A) decrease sample volume

  • 24

    which of the following will improve the lateral resolution on the image?

    C) increase the focal zone number

  • 25

    Transesophageal and transvaginal transducers are considered ___ Devices that should be disinfected using ____

    A) semi critical, high-level disinfectant

  • 26

    how is the pulse repetition period reduced in Doppler imaging

    B) decrease imaging depth

  • 27

    Which of the following will maintain compliance with the ALARA principal

    B) reducing the number of images taken in standard protocol

  • 28

    ____ compares each reflected echo to the same echo from a previous ultrasound pulse to determine if motion is present

    B) autocorrelation

  • 29

    if you increase the sector angle from 45 to 90°, and no other console changes are made what happens to the frame rate

    B) it is reduced to 1/2 the original frame rate

  • 30

    which of the following system controls would increase the color display on the image

    A) image threshold settings

  • 31

    which of the following transducers would be most useful for imaging deeper structures?

    A) 3 MHz long focus

  • 32

    increasing ____ Will increase the ability of the system to display fine details on the image

    C) pixel density

  • 33

    blood flow volume changes in the body are mainly controlled by what two factors

    A) cardiac function and peripheral resistance

  • 34

    in a resting patient the highest resistance in the arterial system is in the

    A) arterioles

  • 35

    The spatial pulse length is 4 mm. what is the minimum axial resolution?

    B) 2 mm

  • 36

    resistance to blood flow is primarily controlled by

    B) the radius of the vessel

  • 37

    if the peak systolic velocity is 3 m/s what is the peak pressure gradient?

    B) 36 mmHg

  • 38

    What component of the beam former protects the sensitivity circuitry of the amplifier

    B) transmit/ receive switch

  • 39

    which of the following will reduce color flash artifact

    A) decrease color priority

  • 40

    what is the best way to reduce sliced thickness artifact that is causing false echoes in a superficial structure

    A) use a standoff pad

  • 41

    what is the cause of the vertical acoustic shadow seen on both images?

    A) element damage

  • 42

    how is detail resolution limit documented on phantom testing?

    D) two separate pins are displayed as a single reflector on the image

  • 43

    The ___ of a wave multiplied by the number of cycles in the pulse will calculate the pulse duration (PD)

    D) period

  • 44

    imaging an adult should never be performed with a thermal index level at or above ___ and imaging A fetus should never be performed with a thermal index level at or above ___

    B) 6,3

  • 45

    while imaging the thyroid you increase the number of focal zones from 1 to 3. How will this affect the image?

    A) degraded temporal resolution

  • 46

    when harmonic imaging is turned on a primarily improves ____ resolution

    D) lateral

  • 47

    which of the following correctly describes how to avoid underestimation of peak velocity measurement on a faint Doppler wave form

    D) increase the gain until speckle artifact is displayed across the screen, then reduce slightly

  • 48

    diagnostic ultrasound is described as frequencies greater than

    B) 2 MHz

  • 49

    for an ultrasound system to produce a real time flicker free image the minimum frame rate required is

    C) 15 Hz

  • 50

    which of the following can change the image data displayed on an archived exam on the ultrasound system

    A) change the gray map used to create the image

  • 51

    review the image in the system, setting along the right side use your mouse and place over the setting that should be increased to improve the diagnostic quality of the color Doppler image

    PRF & WF ( green )

  • 52

    which of the following will lead to more color displayed in real time in the abdominal aorta

    C) increase frame rate

  • 53

    ____ used to form the sector shaped field of view for a phased array transducer

    D) electronic steering is

  • 54

    as the frequency of the wave doubles the wavelength ___

    A) deceased by half

  • 55

    The more bits per pixel the __ shades of gray on the image and ____ improves

    B) more, contrast resolution

  • 56

    if a 7 MHz wave is transmitted through the media listed below the shortest wavelength will be seen in the wave moving through ___

    A) air

  • 57

    ____ is defined as the range of frequencies contained in the U/S pulse

    B) bandwidth

  • 58

    Which of the following is a primary factor in visualization of tissue boundaries that are not perpendicular to the incident beam such as tissue parenchyma and organ boundaries

    B) backscatter

  • 59

    what type of transducer produce the image?

    A) sector phased array

  • 60

    if the dynamic range is 64 dB and there are 32 shades of gray what is the contrast resolution?

    C) 2 dB/shade

  • 61

    in the electronically focused ultrasound beam then intensity is

    B) greatest at the focal point

  • 62

    Which of the following will be negatively affected by spatial compounding

    A) temporal resolution

  • 63

    The ___ is defined as the time it take for one pulse to occur

    B) pulse duration

  • 64

    what happens to the attenuation coefficient as the wave travels through soft tissue

    A) remains constant

  • 65

    Aliasing artifact does not occur with CW Doppler systems because

    D) CW doppler transducers used two separate elements so they are not limited by pulse echo parameters like PRF and Nyquist limit

  • 66

    when you adjust the Doppler PRF you adjust the ___ control on the system

    D) velocity scale

  • 67

    activating harmonic imaging will result in ____

    C) improved contrast resolution

  • 68

    which of the following ultrasound techniques has the highest temporal resolution

    A) M-mode

  • 69

    which of the following describes the best way to improve the doppler tracing of the internal carotid artery

    D) increase the cursor angle to 60°

  • 70

    what acoustic artifact is demonstrated on the image below

    A) posterior shadowing

  • 71

    transducer can be electronically focused by

    C) varying timing of the release of the pulses

  • 72

    the best obtainable angle of insonation for vascular evaluation that is most commonly used to provide Doppler shift information is ___ degrees

    C) 60

  • 73

    which of the following is the best way to reduce the artifactual echoes on the image indicated by the red arrows

    D) reduce TGC settings in the near field

  • 74

    what is the best way to increase beam penetration

    A) decrease operating frequency

  • 75

    ____ is the process that is performed in the beam former where the output voltage is better to decrease the formation of lobe artifacts

    A) apodization

  • 76

    ____ artifact is caused by the refraction and results in subtle image degradation

    D) multipath

  • 77

    increasing the PRF will reduce the appearance of ___ while increasing the appearance of __ on Doppler evaluation

    C) aliasing, range ambiguity

  • 78

    you are performing real time imaging of a vessel adjacent to the tibia what acoustic artifact Will most likely be identified on the 2-D image

    B) posterior shadowing

  • 79

    Which of the following transducers uses a beam former

    B) sector phased array

  • 80

    What setting should be increased to improve the diagnostic quality of color doppler image

    PRF & WF

  • 81

    which of the following system adjustments would increase the size of the spectral window

    D) changing from a 4mm to 2mm sample size

  • 82

    in order to assess low velocity flow which of the following operator adjustments should be made

    D) decrease the rejection and wall filter setting

  • 83

    propagation speed, artifact reverberation artifact and the bayonet artifact are all acoustic artifacts seen with?

    D) needles

  • 84

    which of the following is considered the capacitive component of the circulatory system

    D) veins

  • 85

    ____ artifact appears as multiple parallel equidistant lines that look like steps on the ladder

    C) reverbation

  • 86

    which statement correctly describes a spectral wave form

    D) the presence of a spectral window indicates laminar flow is present in the vessel being evaluated

  • 87

    which of the following is related to an increase in the spatial pulse length of an ultrasound beam produced by the transducer

    B) decease in probe frequency

  • 88

    which of the following adjustments would improve the appearance of ultrasound enhancing agents on the image

    C) decrease output power

  • 89

    how do the microbubbles in an ultrasound contrast agent change the image

    A) increase the reflectivity of flowing blood

  • 90

    what is the proper method used to assess the size of the dead zone?

    B) measure from the top of the image to the depth that uniform tissue is identified

  • 91

    diagnostic ultrasound is described as frequencies greater than

    B) 2 MHz

  • 92

    which of the following describes the effect on the frame rate when changing from duplex to triplex imaging

    C) decreased

  • 93

    Blood flow at a focal stenosis will demonstrate

    B) increased resistance

  • 94

    what change in system controls can reduce the appearance of color displayed over vessel walls

    A) decreasing color priority

  • 95

    what artifact is demonstrated on the ultrasound image by the red arrows?

    D) posterior enhancement

  • 96

    what determines the frequency of a captivated micromachined ultrasound transducer

    D) frequency of the drive voltage

  • 97

    the acoustic power is changed by -3dB the beam intensity ___

    C) decreases by half

  • 98

    which of the pulses produced from a 7MHz transducer would have the best lateral and axial resolution

    D) 2mm beam diameter, 3 pulses per cycle

  • 99

    when working in a hospital setting what is the best way to verify a patient’s identity

    D) check the hospital issued ID bracelet

  • 100

    As the area of the beam increases _______

    A) the intensity of the beam decreases