記憶度
13問
33問
0問
0問
0問
アカウント登録して、解答結果を保存しよう
問題一覧
1
Absorbed dose measurement methods
Calorimetry Fricke dosimetry Film dosimetry Ionization method Thermoluminescence dosimetry
2
is the best technique for absolute and direct measurement of absorbed dose.
calorimetry
3
It is based on the measurement of the temperature rise of a given volume element of this instrument when it is exposed to a source of radiation.
calorimetry
4
also called ferrous sulfate dosimetry, is one of the most useful chemical dosimeters in existence. This dosimetry technique depends on the oxidation of ferrous ions (Fe2+) to ferric ions (Fe3+) by ionizing radiation
fricke dosimetry
5
Fricke dosimetry also called as
ferrous sulfate dosimetry
6
as been applied to measure relative dose distributions in an anthropomorphic polystyrene breast phantom having cork lungs, simulating a radiation therapy treatment with 6 MV opposing tangential beams. ... These results demonstrate that film dosimetry is a useful tool for treatment planning verification.
film dosimetry
7
provides a real advantage to radiation therapy departments in that it allows rapid and easy calibration of output at the end of treatment cones.
thermoluminescence dosimetry
8
Dosimetry Parameters
output factor isodose tpr tar tmr pdd sar smr backscatter factor collimator scatter factor
9
Define as the ratio of the dose rate at the depth maximum dose for the reference field size to the reference field size.
output factor
10
It represents points of equal dose.
isodose
11
It is define as the ratio of the dose at a specific point in tissue or in a phantom to the dose at the same distance in the beam at a reference depth, usually 5 cm.
tpr
12
It is define as the ratio of the dose at a specific point in tissue or in a phantom to the dose at the same distance in the beam at a reference depth, usually HOW MUCH
5 cm
13
It is the ratio of the dose at a given point in a medium to the dose at the same point in the free space
tar
14
Is a special case of TPR where the reference depth is chosen to be at Dmax.
tissue maximum ratio
15
It is the absorbed dose at a given depth expressed as a percentage of the absorbed dose at reference depth on the central axis of the field.
pdd
16
It is the ratio of the scattered dose at a given point in a medium to the dose in air at the same point.
scatter air ratio
17
It is really just a variation of SAR and, like SAR, is mainly used in calculation of scattered dose in phantom or tissue.
scatter maximum ratio
18
Is defined as the ratio of dose on the central axis at Dmax to the dose at the same point in air.
backscatter factor
19
Output measured without a scattering phantom, sometimes referred to as in-air, increases with increase in field size.
collimator scatter factor
20
Typical Field Arrangement
Single Field Parallel opposed field Multiple field Wedge field Moving Beam Treatment
21
Superficial targets
single field
22
Important factors to consider in choosing single field
The range of dose within the tumor The maximum dose in tissues The exit dose
23
Probably the most commonly used field of arrangement.
parallel opposed
24
Often used to treat metastatic disease in various sites.
parallel opposed
25
Frequently employed in treating large fields.
parallel opposed
26
Also often used during a portion of disease of the brain, pelvis, head and neck, lungs, pancreas, and extremities.
parallel opposed
27
Often used when the depth of the target is uncertain.
parallel opposed l
28
The target volume extends throughout a particular area treated.
parallel opposed
29
Multiple angled fields are possible to set up.
parallel opposed
30
More reproducible
parallel opposed
31
Risks of tumor miss are reduced.
parallel opposed
32
The incidental irradiation of large volume of normal tissues situated within the fields
parallel opposed
33
High dose areas can be avoided by using a multiple field arrangement in which all beams encompass the target include different areas of surrounding normal tissues.
multiple field arrangements
34
Certain beam angles are prohibited due to the presence of critical organs in those direction
multiple field arrangements
35
The setup accuracy of a treatment may be better with what?
parallel opposed
36
The acceptability of treatment plan depends not only on the dose distribution but also on:
practical feasibility setup accuracy reproducibility of the treatment
37
Can be designed to produce uniform dose distributions.
multi field arrangements
38
What field is box technique
multiple field arrangements
39
Box technique is for what organ
pelvic tumor
40
Three field technique is for what organs
pancreas bladder rectum
41
The used of these is use to alter the dose distribution in beam.
wedge field
42
Originally designed to be used in the treatment of small shallow tumors..
wedge field
43
Rotation therapy
moving beam
44
Arc Treatment
moving beam
45
Combining Fixed and Moving-beam Technique.
moving beam
46
Treatment preparation
gross tumor clinical target volume planning target volume treated volume irradiated volume
47
The gross palpable or visible/demonstrable extent and location of malignant growth.
gross tumor volume
48
Is a tissue volume a demonstrable GTV and/or subclinical microscopic malignant disease must be eliminated.
clinical target volume
49
Contains the tissues that are to be irradiated to a specific dose according to a specified time-dose pattern.
planning target volume
50
Is the volume enclosed by an isodose surface, selected and specified by radiation oncologist as being appropriate to achieve the purpose of treatment.
treated volume
51
Is the volume of tissue that receive a dose considered significant in relation to tissue tolerance.
irradiated volume
52
Types of beam modification
Shielding Compensation Wedge filtration Flattening
53
Types of beam modification devices
Field blocking and shaping devices Compensators Beam spoilers Wedge filters Beam flattening Bolus Breast cone Penumbra trimmers Electron beam modification
54
Field blocking shaping filters
Shielding blocks Custom blocks Asymmetrical jaws Multi- leaf collimators
55
Refers to the material whose interactions with radiation beam mimic those tissue.
bolus
56
Bolus devices
Paraffin wax Vaseline gauze Wet gauze or towels Water bags
57
Bolus of a thickness equal to the depth of maximum dose eliminates WHAT of megavoltage photon beams.
skin sparing effect
58
It can also be used to compensate variations in contour surface contour to eliminate air gaps in cavities.
bolus
59
Skewed dose distribution.
compensator
60
Placed in the head of treatment unit, thus modifying the radiation beam to accommodate the contour of the patient.
compensator
61
Compensator devices
brass copper lead lucite
62
The dimension and shape of the compensator must be adjusted to account for
Beam divergenc Linear attenuation coefficient of the filter material and soft tissue Reduction in scatter
63
As the distance between the skin and compensator increase the thickness ratio what?
decreases
64
As the distance between the skin and compensator what? the thickness ratio decreases
increases
65
Compensator consideration
Compensator to surface distance (less than 20 cm) Field size Depth Beam quality
66
It is designed to change the isodose curve relative to the beam axis at specified depth to the patient. Wedge reduce the dose in areas of overlap between fields that have hinge angles. Less than 180 degrees.
wedges
67
Sizes of wedges
15, 30, 45 and 60 degrees
68
Wedges are made of
Tungsten, lead, steel
69
Wedge Placed how much from the skin surface
15 cm
70
Wedge Placed 15 cm from what
skin surface
71
Individual wedges - used in what?
cobalt
72
used in cobalt
individual wedge
73
Wedges
Individual wedges Universal wedges Dynamic wedges Virtual wedges Pseudo wedges
74
In wedge x and y is what
width, legth
75
Special beam modifiers where shadow trays made from lucite are kept at a certain distance from the skin
beam spoilers
76
Allow to transmit of predetermined percentage of the treatment beam to a portion of the treatment field and may be used throughout the course of the treatment.
transmission filter
77
Used to shape photon or electron fields, take several forms. Materials range from spent uranium to lead to lead alloys.
shielding blocks
78
Used to shape photon or electron fields, take several forms. Materials range from spent what
uranium to lead to lead alloys.
79
Full shielding blocks are constructed to transmit how much of the original beam
less than 5%
80
Cerrobend (lipowitz metal) melting point is what?
165 degrees F (74 degrees Celsius)
81
Its melting point 165 degrees F (74 degrees Celsius)
Cerrobend (lipowitz metal)
82
Bismuth transmit how many percentage
50
83
Leadtransmit how many percentage
26.7
84
Tin transmit how many percentage
13.3
85
Cadnium transmit how many percentage
10
86
Shielding blocks thickness
7.5 cm
87
Shielding blocks 9.4 g/cm3 at how many degrees
20
88
Shielding blocks how much at 20 degrees
9.4 g/cm3
89
Two types of shielding blocks
positive, negative