暗記メーカー
ログイン
CLUSTER 2
  • GAJE, Harvey Jay M.

  • 問題数 100 • 11/26/2023

    記憶度

    完璧

    15

    覚えた

    35

    うろ覚え

    0

    苦手

    0

    未解答

    0

    アカウント登録して、解答結果を保存しよう

    問題一覧

  • 1

    Rare earth screens are most effective in reducing patient dose in

    abdominal radiography

  • 2

    Of the following elements, which has the highest atomic number?

    tungsten

  • 3

    Rare earth phosphors have conversion efficiencies ______Those of calcium tungstate

    3-4 times

  • 4

    Which of the following spectra is discrete?

    rare earth emission

  • 5

    Rare earth screens

    Show higher x-ray absorption in the diagnostic range than calcium tungstate screens

  • 6

    When diagnostic x-rays interact with rare earth phosphors,

    The lower the x-ray energy the more probable is photoelectric absorption

  • 7

    When using rare earth intensifying screens.

    The abrupt change in absorption at an energy equal to the K-shell binding energy called the K-shell absorption edge.

  • 8

    By comparison with calcium tungstate, when an x-ray is absorbed in an rare earth screen

    the more light will be emitted

  • 9

    In comparison with calcium tungstate screens, rare earth screens

    absorb three to five times as many x-rays

  • 10

    Rare earth intensifying screens

    Are more expensive than calcium tungstate screens

  • 11

    Which of the following statements concerning intensifying screens is correct?

    All have conversion efficiencies greater than one

  • 12

    What is the principal limitation of the rare earth intensifying screen?

    quantum mottle

  • 13

    Stains in intensifying screens from processing chemicals are likely to result in ______ in the radiograph

    light blotches

  • 14

    In the proper design and use of a film cassette.

    The front cover should be made of low Z material to minimize x-ray attenuation

  • 15

    Carbon fiber is particularly useful in x-ray imaging because of its

    lower atomic number

  • 16

    The routine care of screens requires that they be

    periodically cleaned

  • 17

    Remnant x-ray include those that

    exit the patient toward the image receptor

  • 18

    Which of the following factors can both affect the level of scatter radiation and be ready controlled by the radiologic Technologist?

    inherent filtration field size

  • 19

    As the kVp of operation increases the absolute level of scatter radiation will

    decrease because of less compton interaction

  • 20

    At 100kVp what is the approximate percent of x-rays that are transmitted through 100m of soft tissue?

    5%

  • 21

    When kVp is increased with a compensating reduction in mAs, then _____ is reduced

    patient dose

  • 22

    If a constant radiographic density is maintained while increasing KVp

    patient dose will increase

  • 23

    As the field size is increased, scatter radiation

    is increased

  • 24

    One factor that influences the production of scatter radiation that is generally not under the control of the technologist is

    patient thickness

  • 25

    Which of the following is not a device designed to reduce the level of scatter radiation Reaching the film?

    a test pattern

  • 26

    Scatter radiation chiefly influences radiographic quality by reducing

    contrast

  • 27

    The x-ray photons that the technologist would like to have interact with the image receptor Are those that are _______ the body

    transmitted through

  • 28

    as kVp is increased from 70-80,

    There will be a higher proportion of scatter radiation

  • 29

    In conventional radiography, most of the photons that reach the image receptor are

    compton-scattered photons

  • 30

    Remnant radiation consist of all emerging from the patient in the direction of the Image receptor.

    photons

  • 31

    As kVp increases from 70-90, all other factors remaining constant.

    quantity of absorbed photons will decrease

  • 32

    As the field size of the x-ray beam increases the ______ increases

    patient exposure

  • 33

    In general, as the thickness of the part being radiographed increases.

    the patient exposure increases

  • 34

    Remnant x-rays include those transmitted and ________ By the patient

    scattered forward

  • 35

    In radiography, one would like to employ very low KVp because

    scattered radiation would be low

  • 36

    In fact, it is necessary to use higher kVp in order that ___________ be at an acceptable level.

    patient dose

  • 37

    In general, if an increase in radiographic technique is required because of patient size, patient dose will be lowest if

    only kVp is increase

  • 38

    X-ray beam restriction to the area of interest is principally required to reduce

    patient dose

  • 39

    Which of the following is not a beam-restricting device?

    added filtration

  • 40

    An aperture diaphragm should allow x-rays to expose an area

    just smaller than the image receptor

  • 41

    An aperture diaphragm is used

    x-ray field cutoff can occur if the diaphragm is not properly positioned.

  • 42

    Dental radiographic cones are usually designed for either ________or _______ source-to- image receptor distances(SID).

    20cm,40cm

  • 43

    The least desirable dental cone is

    short,plastic pointer cone

  • 44

    Off-focus radiation

    degrade the image receptor

  • 45

    In a light-localizing, variable-aperture collimator,

    Periodic checks of x-ray beam and light-field coincidence are necessary

  • 46

    The simplest of all beam restricting devices is

    aperture diaphragm

  • 47

    Which of the following devices is normally designed to limit off-focus radiation?

    First stage shutters of a variable of a variable-aperture collimator

  • 48

    Which of the following could be considered a beam-resisting device?

    fluoroscopic shutter

  • 49

    Cone cutting

    Occurs if the axis if the cone, tube and image receptor are not aligned

  • 50

    if a fixed aperture, rectangular, beam – restricting device is used, then

    An unexposed border should be visible on all four sides of the radiograph

  • 51

    A properly designed light-localizing, variable-aperture collimator will

    be design to reduce off-focus radiation

  • 52

    Radiographic cones and cylinder are employed principally to reduce

    patient dose

  • 53

    in dental radiography

    the x-ray beam diameter should not exceed 7cm

  • 54

    Which of the following is most useful in dental radiography?

    rectangular aperture diaphragm

  • 55

    PBL stands for

    positive beam limiting

  • 56

    One helpful feature of a light localizing collimator is increased

    filtration

  • 57

    A diaphragm is machined to just match image receptor size. If an unexposed border is required on the radiograph, the diaphragm opening will

    have to be reduced

  • 58

    A dental cone diaphragm assembly must restrict the circular x-ray beam to 7cm diameter at the tip of the cone. If the source-to-cone tip distance is 20cm and the source-16- Diaphragm distance is 4in what should the diameter of the diaphragm be?

    1.4 cm

  • 59

    An aperture diaphragm is designed for 10x12 in film. If the SID is 36 inches and the source-to-diaphragm distance is 4cm, what should the diameter of the diaphragm be?

    in x 1.3 in

  • 60

    A room to be used exclusively to produce 14 in x17 in chest radiographs is being installed The SID is 180 cm and the source-to-diaphragm distance is 10cm, What size should be the opening of the diaphragm be?

    0.8 cm x 0.9 in

  • 61

    The principal reason for using a grid is to

    increase radiographic contrast

  • 62

    Which of the following will not enhance radiographic contrast?

    the use of added filtration

  • 63

    If one had to select the single most important grid characteristics, it would be

    grid ratio

  • 64

    In a grid that has lead strips 0.5mm apart and 4mm high, the grid ratio is

    8:1

  • 65

    In a grid that has lead strips 0 5mm apart and 8mm high, the grid ratio is

    16:1

  • 66

    The use of a grid will result in

    control scatter radiation

  • 67

    The design of a radiographic grid, the

    interpace material is radiolucent

  • 68

    X-ray grids are principally effective in attenuating

    x-rays following compton interaction

  • 69

    If the interspace dimension is constant, increasing the grid ratio will.

    make grid thicker

  • 70

    As grid frequency increases,

    The interspace width becomes thinner if the width of the grid, remains constant

  • 71

    Grid with a high ratio

    are more effective than those with low ratio

  • 72

    The efficiency of a grid for reducing scatter radiation is related principally to the

    grid ratio

  • 73

    The patient dose increases as

    as the atomic number of interspace material increase

  • 74

    Radiographic grids

    must be place between the patient and the image receptor

  • 75

    Added filtration influences selection of

    high energy x-rays

  • 76

    If only scatter radiation reached the image receptor

    radiographic contrast would be very low

  • 77

    A grid has the following characteristics grid ratio – 10:1. Grid strip = 25um, interspace width =300um, contrast improvement factor = 1.8 What percentage of the incident radiation will be intercepted?

    7.7%

  • 78

    A grid has the following characteristics grid frequency = 40lines/cm, grid strip width= 30um, grid height 3.0mm. The grid ratio is approximately

    14:1

  • 79

    A grid has the following characteristics grid ratio=81 Grid strip width=50um, grid height -3.2mm. The grid frequency is approximately

    22 lines/cm

  • 80

    Which of the following is the least important indicator of a grid performance?

    grid strip height

  • 81

    The contrast improvement factor is defined a the radiographic contrast obtained.

    with a grid compared to that obtained without one

  • 82

    383. As the ratio of grids increases, The __________also increases

    contrast improvement factor

  • 83

    The use of high-frequency grids result in

    increased contrast improvement factor

  • 84

    Radiographic grids with high contrast improvement usually

    increased radiographic contrast

  • 85

    A radiograph is made at 76kVp 25 mAs without a grid. If an 8:1 ratio grid is added, the mAs required would the be approximately

    75 mAs

  • 86

    The simplest type of grid is the

    linear grid

  • 87

    The undesirable absorption of primary-beam x-rays by a grid is called

    grid cut off

  • 88

    Which of the following would be included in the three major classifications of moving grids?

    reciprocating

  • 89

    If one had two grids whose characteristics were unknown, but grid B weighed twice as Much as grid A, one might conclude that grid B would have a

    higher contrast improvement factor

  • 90

    Focused grids

    Cut off two edges of an image if placed too far from the tube

  • 91

    One factor that does not affect the percentage of scatter radiation reaching the image receptor is

    mAs

  • 92

    The efficiency of a grid for scatter radiation cleanup is related principally to the grid

    ratio

  • 93

    Radiographic grid

    Usually have grid ratios between 5:1 and 16:1

  • 94

    The contrast improvement factor

    varies with the kVp of operation

  • 95

    Generally the selectivity of a grid will depend principally on

    grid mass

  • 96

    Which the moving-grid technique, the appearance of grid lines could indicate that before the grid moved

    the exposure started

  • 97

    The average gradient of an image receptor is 2 1 With a 10:1 grid having 50lines/cm frequency, the average gradient is increased to 3.2 What is the contrast improvement factor?

    1.5

  • 98

    Bucky factor is a measure of

    penetrability of a beam through a grid

  • 99

    The value of the Bucky factor increases with

    increasing x-ray quality

  • 100

    Which of the following is not a grid positioning error?

    air-gap grid