暗記メーカー
ログイン
M1
  • E

  • 問題数 37 • 4/18/2025

    記憶度

    完璧

    5

    覚えた

    15

    うろ覚え

    0

    苦手

    0

    未解答

    0

    アカウント登録して、解答結果を保存しよう

    問題一覧

  • 1

    graphic representation of a data set

    Histogram

  • 2

    includes all the pixel values represent the image before edge detection and rescaling.

    Data set

  • 3

    6 Histogram Analysis Errors IEMULI

    - Incorrect Anatomic Menu Selection - EXPOSURE FIELD RECOGNITION ERROR - Multiple fields/plate - UNEXPECTED MATERIAL IN DATA SET - LARGE OVEREXPOSURE ERROR - INAPPROPRIATE RESCALING

  • 4

    should be selected prior to processing the imaging plate; otherwise the image will not process the correct density and contrast.

    Correct anatomical menu

  • 5

    If the x-ray exposure field is not properly collimated, sized and positioned, this error would occur.

    Exposure field recognition error

  • 6

    •For image histogram to be properly analyzed each collimated field should consist of how many distinct collimated images?

    Four

  • 7

    must not overlap and must be parallel to the edges of the plate, otherwise the image may be processed incorrectly and the exposure indicators will be incorrect

    Collimated borders

  • 8

    If the exposure pattern does not match one of the segments patterns, the data from the whole plate is included in the?

    histogram

  • 9

    Is single plate worse?

    Better

  • 10

    If there is unexpected material in data set, image processing and exposure indicators may be?

    Incorrect

  • 11

    During large overexposure error, Histogram will try to correct the image but will remove what to do it

    Information

  • 12

    Under inappropriate rescaling, If the histogram has excess exposure from poor collimation techniques, this could cause an error in locating the ____

    Value of Interest (VOI)

  • 13

    •The imaging plate of the Computed radiography system is sensitive to these two types of radiation

    - Background radiation - Scattered radiation

  • 14

    makes scatter control much more important with a digital system (CR) when compared with screen/ film

    Low Exposure Response and Dynamic Range

  • 15

    • PSPs respond to exposures as low as

    60uR

  • 16

    Background radiation measures

    40uR - 80uR/day.

  • 17

    more sensitive to lower-energy levels of radiation such as scatter, which results in increased fog in the image.

    Digital IRs

  • 18

    Two major factors that affects the amount and energy of scatter radiation exiting the patient: KV

    • 1. kilovoltage peak (kVp) and • 2. volume of tissue irradiated.

  • 19

    this is called the limiting of patient exposure and reducing the amount of scatter radiation produced within the patient.

    Beam limitation

  • 20

    - A device that has very thin lead strips with radiolucent interspaces, intended to absorb scatter radiation emitted from the patient. - placed between the patient and the IR, are invaluable in the practice of radiography. - absorb, or clean up, all scattered photons directed toward the IR and would allow all transmitted photons emitted from the patient to pass from the patient to the IR.

    Grid

  • 21

    can be described by grid frequency and grid ratio

    Grid construction

  • 22

    the # of gridlines per unit distance.

    Grid frequency

  • 23

    can range in value from 25 to 80lines/cm (63 to 200 lines/inch).

    Grid frequencies

  • 24

    defined as the ratio of the height of the lead strips to the distance between them.

    Grid ratio

  • 25

    Grids are typically used only when the anatomic part is __ cm or greater in thickness, and more than __ kVp is needed for the examination

    - 10 cm - 60 kVp

  • 26

    The _________ is scanned line by line; IF the SCAN FREQUENCY and the ___ FREQUENCY are similar and oriented in the same direction, a _______ effect will be observed.

    - imaging plate - Grid Frequency - Moire Effect

  • 27

    This refers to the linear pattern of the lead lines of a grid.

    Grid Pattern

  • 28

    Two Grid Patterns LC

    1. Linear 2. Crosshatched/Crossed

  • 29

    has lead strips running parallel to the long axis of the grid.

    • •Long Dimension Grids-

  • 30

    has lead strips running perpendicular to the long axis of the grid. may be useful for examinations where it is difficult to center the central ray correctly

    Short dimension grid

  • 31

    4 Grid Types PCFM

    • Parallel Grid • Crossed Grid • Focused Grid • Moving Grid

  • 32

    Refers to a decrease in the number of transmitted photons that reach the IR because of some misalignment of the grid. Further reduction in the number of photons reaching the IR.

    Grid Cut Off

  • 33

    6 TYPES OF GRID CUT OFF ERRORS UFOOO

    • Upside-down • Focused • Off level • Off center • Off focus

  • 34

    • grid cut off occurs when a focused grid is placed in this direction on the IR, resulting in the grid lines going opposite the angle of divergence of the x-ray beam. • Significant loss of exposure along the edges of the image • easily avoided because every grid should have a label indicating "Tube Side”

    Upside down error

  • 35

    - Results when the x-ray beam is angled across the lead strips. It is the most common type of cutoff and can occur from either the tube or the grid being angled. - seen with mobile radiographic studies or horizontal beam examinations and appears as a loss of exposure across the entire image.

    Off level error

  • 36

    Also called lateral decentering, occurs when the central ray of the x-ray beam is not aligned from side to side with the center of a focused grid. - appears as an overall loss of density on radiographic film.

    Off center error

  • 37

    - occurs when using an SID outside of the recommended focal range. - occurs if the SID is less than or greater than the focal range. - It appears the same radiographically as a loss of density at the periphery of the film.

    Off focus grid cutoff error