CHAPTER 8
問題一覧
1
sterilization
2
drain line
3
over the drain
4
steam under pressure, time and temperature
5
10 minutes
6
immediate use steam sterilizer
7
drying
8
18 mins, 273F (134C), Prevacid cycle
9
steam
10
gravity steam
11
reprocess the item
12
superheating
13
yellow
14
on the top rack
15
basin sets
16
the manufacturer provides technical data showing that it is acceptable
17
Bowie dick test
18
biological
19
draw a line through the item and initial
20
chemical
21
class 5
22
geobacillus stearithermophilus
23
gravity displacement liquid cycle
24
3 to 3 1/2 mins at 270F (132C) dynamic air removal
25
once a day and if a new lot number is used
26
implantable devices
27
as recommended by the manufacturer of the container
28
within the first hour after sterilization
29
clean hands
30
consult the manufacturers written instructions for approved sterilization methods and parameters
31
chamber drain line
32
dynamic air removal and gravity displacement
33
100% ethylene oxide
34
bacterial cells
35
exposure
36
alkylation
37
geometric center of the load
38
sterilizer is able to sterilize the contents
39
before the Bowie dick tests
40
on a regular basis according to M.I
41
time,temperature,and hydrogen peroxide gas plasma
42
geobacillus stearothermophilus
43
113F-131F
44
geobacillus stearothermophilus
45
dry heat
46
bacillus atrophaeus
47
100000
48
131F-140F
49
all loads back to the last negative BI
50
condensation
51
did not kill all the test spores
52
1 inch
53
total hip replacement
54
12 hours @125F
55
4 mins, 270F, 28-30psi
56
none of the above
57
all the above
58
25lbs
59
release the load for use
60
59%
61
print outs
62
2 chemical indicators
63
LTHP
64
human error
65
class 4
66
because air tends concentrate at the bottom of the chamber
67
reduce the amount of condensate that forms
68
ozone
69
12
70
60-80psi
71
bacillus atrophaeus
72
perform in parallel to a BI
73
dry heat
74
3 1/2 -4 mins, 270F pre vac
75
30-120mins
76
at the end of each cycle
77
all plastic pouches
78
from red to yellow
79
LTGP
80
all of the above
81
exposure time
82
remove from the sterilizer cart
83
time, temp, and hydrogen peroxide concentration
84
sterilizer jacket
85
ethylene oxide
86
97%-100%
87
latex gloves
88
a mixture of metal and plastics
89
condensation and moisture
90
all of the above
91
monthly
92
all of the above
93
diffusion
94
purple
95
ozone
96
drain line basket and the door gasket
97
master product
98
plasma
99
CDC
100
99
問題一覧
1
sterilization
2
drain line
3
over the drain
4
steam under pressure, time and temperature
5
10 minutes
6
immediate use steam sterilizer
7
drying
8
18 mins, 273F (134C), Prevacid cycle
9
steam
10
gravity steam
11
reprocess the item
12
superheating
13
yellow
14
on the top rack
15
basin sets
16
the manufacturer provides technical data showing that it is acceptable
17
Bowie dick test
18
biological
19
draw a line through the item and initial
20
chemical
21
class 5
22
geobacillus stearithermophilus
23
gravity displacement liquid cycle
24
3 to 3 1/2 mins at 270F (132C) dynamic air removal
25
once a day and if a new lot number is used
26
implantable devices
27
as recommended by the manufacturer of the container
28
within the first hour after sterilization
29
clean hands
30
consult the manufacturers written instructions for approved sterilization methods and parameters
31
chamber drain line
32
dynamic air removal and gravity displacement
33
100% ethylene oxide
34
bacterial cells
35
exposure
36
alkylation
37
geometric center of the load
38
sterilizer is able to sterilize the contents
39
before the Bowie dick tests
40
on a regular basis according to M.I
41
time,temperature,and hydrogen peroxide gas plasma
42
geobacillus stearothermophilus
43
113F-131F
44
geobacillus stearothermophilus
45
dry heat
46
bacillus atrophaeus
47
100000
48
131F-140F
49
all loads back to the last negative BI
50
condensation
51
did not kill all the test spores
52
1 inch
53
total hip replacement
54
12 hours @125F
55
4 mins, 270F, 28-30psi
56
none of the above
57
all the above
58
25lbs
59
release the load for use
60
59%
61
print outs
62
2 chemical indicators
63
LTHP
64
human error
65
class 4
66
because air tends concentrate at the bottom of the chamber
67
reduce the amount of condensate that forms
68
ozone
69
12
70
60-80psi
71
bacillus atrophaeus
72
perform in parallel to a BI
73
dry heat
74
3 1/2 -4 mins, 270F pre vac
75
30-120mins
76
at the end of each cycle
77
all plastic pouches
78
from red to yellow
79
LTGP
80
all of the above
81
exposure time
82
remove from the sterilizer cart
83
time, temp, and hydrogen peroxide concentration
84
sterilizer jacket
85
ethylene oxide
86
97%-100%
87
latex gloves
88
a mixture of metal and plastics
89
condensation and moisture
90
all of the above
91
monthly
92
all of the above
93
diffusion
94
purple
95
ozone
96
drain line basket and the door gasket
97
master product
98
plasma
99
CDC
100
99